Infant Incubator Use Guide

The baby incubator is an advanced technology that integrates various disciplines such as clinical medicine, machinery, computer automatic control, and sensors. It provides a prenatal and sick infant with an air purification temperature and humidity suitable, and an excellent environment similar to the maternal uterus. The space temperature and humidity can be set according to the doctor's orders. The baby's body temperature, space temperature, and space humidity are all digitally displayed, exceeding the normal value, with sound and light alarms to ensure the reliability and safety of the machine.
1. Operational matters needing attention
1. Before using the baby incubator, please read and understand the contents of the manual.
2. Improper operation of the incubator will cause harm to the baby. Therefore, we should be familiar with the use of incubators and the generally known risks and benefits, and we must pass special training to operate them.
3. The incubator must be set to a temperature value after being turned on and used after 60 minutes of preheating.
4. If the incubator issues a "power failure alarm", "fan alarm" or "over temperature alarm" warning message, please do not use the incubator. Relevant maintenance services should be performed by qualified personnel.
5. Before opening the incubator thermostat, make sure to cut off the wiring of the sensor part.
6. Do not open the incubator thermostat. While the baby is nursing in the incubator, there is no need to open the incubator thermostat at any time. All necessary contact with the baby can be done through the front door baffle and operating window.
7. If the front door baffle is opened, the temperature displayed in the temperature display window may not be the actual temperature in the incubator. Therefore, do not leave the front main door open for a long time.
8. All baffle latches should be firmly inserted to prevent the phenomenon of opening outside the baffle.
9. To ensure the safety of the baby, do not leave the baby in the incubator unattended while the front door baffle is open. Do not open the incubator thermostat when lifting the crib.
10. The use of other auxiliary equipment in the incubator may cause changes in the air flow pattern, and may also affect the consistency and variability of the temperature, and the relative relationship between the temperature reading of the incubator and the temperature of the baby's skin surface .
11. If debris (such as blankets, animal toys, etc.) in the air circulation channel causes obstruction, it will affect the safety of the baby and the performance of the incubator during clinical use.
12. In order to prevent direct radiation from causing overheating of the baby, do not place the incubator in direct sunlight or receive other sources of radiant heat.
13. Do not put blankets or the like around the baby or at the vents. This may cause the baby to overheat, resulting in injury or burns.
14. If the light treatment equipment is placed too close to the incubator thermostat, it will affect the temperature of the incubator thermostat, the temperature in the incubator and the temperature of the baby's skin surface.
15. In order to operate the incubator correctly, only the skin temperature sensor provided by our company can be used.
16. Never put the skin temperature sensor on the back of the baby or in their rectum.
17. In the skin temperature control mode, the probe of the skin temperature sensor must directly contact the surface of the baby's skin to provide accurate temperature control information of the surface temperature of the baby's skin. In this state, if the sensor probe is not in direct contact with the baby's skin, it may cause the baby's skin to overheat. The temperature of the baby ’s body temperature must be checked regularly to ensure that the sensor is in the correct contact state and the baby ’s skin is overheated.
18. The incubator must be fixed on a base by the locking device provided by itself. If it is not fixed, when the incubator is tilted, especially when the incubator thermostat is opened, the incubator will detach from the base.
19. The leakage of current flowing through all equipment in the incubator, including the device in the socket part, shall not exceed 500 microamperes.
20 To prevent the incubator from hitting or damaging nearby carts, shelves or tables, these items must be placed at least 30 cm away from the incubator.
21. In order to prevent the incubator from sliding down when stopped on the slope, the locking casters on the front of the base must be down and locked along the slope.
22. In order to prevent accidental disconnection, when adjusting the height of the crib, make sure that there is sufficient length between each wire, infusion tube and ventilation pipe to the crib mattress, so that the crib height can be adjusted in all directions.
23. When moving the incubator, in order to prevent injury or damage, at least one person with corresponding strength must perform corresponding control.
24. In order to maintain the best stability of the incubator during the movement, before moving, make sure that all items are firmly fixed in the position.
25. The use of auxiliary equipment that meets the safety requirements of the equipment will reduce the safety performance of the system. The following points must be considered when selecting relevant equipment:
(1) Use auxiliary equipment in the patient area.
(2) Evidence shows that the auxiliary equipment has been tested for safety in accordance with relevant national standards and obtained a safety certificate.
26. The incubator has no oxygen delivery device and humidity control device. If you need to increase the humidity in the incubator, please add distilled water to the cup.
27. When the power of the whole machine is not connected, the power switch should not be turned on for a long time, otherwise the power failure alarm will occur and the battery will be wasted.
28. The temperature display module of the temperature controller is susceptible to interference from electromagnetic fields, so it should not be used in high electromagnetic field strength situations.
29. When the exterior of the temperature controller is an F-type 3A fuse tube, it must be replaced with a fuse tube of the same specification, otherwise the safety performance of the system will be reduced.
30. When the instrument is working, medical personnel must regularly (at least once every 30 minutes) check the baby's condition and monitor the baby's temperature to see if there is overheating or cold.
31. If it is used at an ambient temperature lower than 20 degrees Celsius, water mist may appear in the incubator, but it does not affect the use effect.